Densitometric functions
............................
4.4.1
Position the measuring aperture on the color samples Black, Cyan, Magenta and Yellow, and initiate
the measurements by pressing the <Measurement key>.
• • • • •
Select the entry line to be corrected by turning the <Control ball> and confirm the selection by
pressing the <Measurement key>.
The entry field for the density reference value appears on the display.
Correct the density reference value by turning the <Control ball> and terminate the entry by
pressing the <Measurement key>.
The entry field for the density tolerance value appears on the display.
Correct the density tolerance value by turning the <Control ball> and terminate the entry by pressing
the <Measurement key>.
Select the menu line Accept data by turning the <Control ball>.
Accept the reference and tolerance values by pressing the <Measurement key>.
Select the menu line Measure paper white by turning the <Control ball>.
Position the measuring aperture on the paper white and initiate a measurement by pressing the
<Measurement key>.
The display White measurement done is shown for about 1 second.