Densitometric functions
............................
4
4.5.1
Select the menu line Accept data by turning the <Control ball>.
Accept the reference and tolerance values by pressing the <Measurement key>.
Select the menu line Measure paper white by turning the <Control ball>.
Position the measuring aperture on the paper white and initiate a measurement by pressing the
<Measurement key>.
The display White measurement done is shown for about 1 second.
Position the measuring aperture on the color sample and initiate a measurement by pressing the
<Measurement key>.
The measured density differences D appear on the display.
If one or more preset tolerance values for the density difference are exceeded, the note Out of
tolerance will additionally be shown.
Select the menu line Return by turning the <Control ball> and actuate the <Measurement key> to
return to main menu.